Dairy Cow Progesterone (P) ELISA Kit
user's Guide
This kit is for research use only.
Detection range: 96T
60 pmol/L -1600pmol / L
purpose of usage:
This kit is used to determine the content of progesterone (P) in cow serum, plasma and related liquid samples. Experimental principle
This kit uses the double antibody sandwich method to determine the level of progesterone (P) in dairy cows. The microtiter plate is coated with purified cow progesterone (P) antibody to make solid phase antibody. Progesterone (P) is added to the monoclonal antibody-coated microwells in turn, and then combined with HRP-labeled progesterone (P) antibody To form an antibody-antigen-enzyme-labeled antibody complex, and after thorough washing, the substrate TMB is added for color development. TMB is converted into blue under the catalysis of HRP enzyme, and into the final yellow under the action of acid. The color depth is positively correlated with the progesterone (P) in the sample. The absorbance (OD value) was measured with a microplate reader at a wavelength of 450nm, and the concentration of progesterone (P) in the cow was calculated from the standard curve. Kit composition
1
30 times concentrated washing solution
20ml × 1 bottle
7
Stop solution
6ml × 1 bottle
2
Enzyme reagent
6ml × 1 bottle
8
Standard product (3200pmol / L)
0.5ml × 1 bottle
3
Enzyme coated plate
12 holes × 8 strips
9
Standard dilution
1.5ml × 1 bottle
4
Sample diluent
6ml × 1 bottle
10
Instructions
1 serving
5
Developer A liquid
6ml × 1 bottle
11
Sealing film
2 sheets
6
Developer B liquid
6ml × 1 / bottle
12
sealed bag
1
Specimen requirements
1. Specimens are extracted as soon as possible after collection, and extraction is performed according to relevant literature, and experiments should be conducted as soon as possible after extraction. If the test cannot be performed immediately, the specimen can be stored at -20 ℃, but repeated freezing and thawing should be avoided
2. The samples containing NaN3 cannot be detected because NaN3 inhibits horseradish peroxidase (HRP) activity. Steps
1. Dilution of standard products: This kit provides one original standard product. The user can perform dilution in a small test tube according to the following chart.
1600pmol / L
Standard No. 5
150µl of original standard is added to 150µl of standard diluent
800pmol / L
Standard 4
Add 150µl of Standard No. 5 to 150µl of Standard Diluent
400pmol / L
Standard 3
150µl Standard No. 4 is added to 150µl Standard Diluent
200pmol / L
Standard No. 2
150µl Standard No. 3 is added to 150µl Standard Diluent
100pmol / L
Standard 1
150µl Standard No. 2 is added to 150µl Standard Diluent
2. Sample addition: set up blank wells (without adding samples and enzyme-labeled reagents in the blank control wells, the rest of the steps are the same), standard wells, and sample wells to be tested Accurately add 50µl of the standard on the enzyme-coated plate, add 40µl of sample diluent to the sample well, and then add 10µl of the sample to be tested (the final dilution of the sample is 5 times). Add the sample and add the sample to the bottom of the well of the microplate, try not to touch the wall of the well, shake gently to mix.
3. Incubation: seal the plate with a sealing film and incubate at 37 ° C for 30 minutes.
4. Mixing solution: Dilute 30 times concentrated washing liquid with distilled water 30 times and reserve
5. Washing: Carefully peel off the sealing film, discard the liquid, spin dry, fill each well with the washing solution, let it stand for 30 seconds and then discard, repeat this 5 times, pat dry.
6. Add enzyme: Add 50μl of enzyme label reagent to each well, except for blank wells.
7. Incubation: The operation is the same as 3.
8. Washing: The operation is the same as 5.
9. Color development: add developer A50µl first to each well, then add developer B50µl, gently shake to mix, and develop color at 37 ° C in the dark for 15 minutes.
10. Termination: Add 50µl of stop solution to each well to stop the reaction (at this time, the blue color turns to yellow).
11. Determination: Measure the absorbance (OD value) of each well in sequence with blank air conditioner zero and 450nm wavelength. Determination should be terminated
Within 15 minutes after the solution. Summary of operating procedures:
Calculation
Taking the concentration of the standard as the abscissa and the OD value as the ordinate, draw a standard curve on the coordinate paper, and find the corresponding concentration from the standard curve according to the OD value of the sample; then multiply it by the dilution factor; Calculate the linear regression equation of the standard curve with the OD value, substitute the OD value of the sample into the equation, calculate the sample concentration, and multiply it by the dilution factor to obtain the actual concentration of the sample. Precautions
1. The kit should be equilibrated at room temperature for 15-30 minutes before being taken out of the refrigerated environment. If the enzyme label coated plate is unopened, the strip should be stored in a sealed bag.
2. Crystals may be precipitated in the concentrated washing liquid, which can be heated and dissolved in a water bath during dilution, and the results will not be affected during washing.
3. The sampler should be used at each step of sample addition, and the accuracy should be regularly checked to avoid test errors. It is best to control the sampling time within 5 minutes. If there are many specimens, it is recommended to use a volley gun to add samples.
4. Please make a standard curve at the same time of each measurement, it is best to make a double hole. If the content of the test substance in the sample is too high (the OD value of the sample is greater than the OD value of the first well of the standard well), please dilute it with a certain multiple (n times) of the sample diluent and then determine it. Multiple (× n × 5).
5. The sealing film is limited to one-time use to avoid cross-contamination.
6. Please keep the substrate away from light.
7. Strictly follow the instructions, and the test results must be determined by the microplate reader.
8. All samples, washing liquids and various wastes should be treated as infectious agents.
9. The components of different batches of this reagent shall not be mixed. 10. If there is any difference with the English manual, the English manual shall prevail. Storage conditions and validity period
1. Kit storage :; 2-8 ℃.
2. Validity: 6 months
Bowl chopper and vacuum mixer are widely used in sausage production line,including bowl cutter,bowl chopper,meat mixer,vacuum bowl cutter,vacuum cutter,vacuum mixer. Only the excellent performance machines can guarantee the high quality of products. 30 years of experience allows us to design the most advanced chopping and mixing machines. Vacuum sealing technology is applied in these machines preventing protein and fat being oxygenized and damaged, keeping freshness, original color and taste, reducing air bubbles and enhancing flexibility.
The vacuum mixer adopts double shaft and helix paddle structure, which will mix raw material in circle movement, to reach the perfect mixing effect.
Bowl cutter can cut meat into emulsified paste by high speed chopping knives and mix up other additives into meat sufficiently.
Bowl cutter series:
High speed bowl cutter 80L
High speed bowl cutter 125L
High speed bowl cutter 200L
High speed bowl cutter 330L
Vacuum bowl cutter 125L
Vacuum bowl cutter 200L
Mixer series
Common meat horizontal mixer 60L, 150L, 400L, 650L, 800L, 1200L
Vacuum meat horizontal mixer 60L, 150L, 300L, 650L, 800L, 1200L
The application of Chopper And Mixer Series
Raw sausage
Cooked sausage
Smoked sausage
Scalded sausage
Canned meat
Pies
Soups and sauces
And much more ...
Industrial Meat Mixer,Vacuum Meat Mixer,Chopper And Mixer Series,Bowl Cutter,Bowl Chopper,Meat Mixer,Vacuum Bowl Cutter,Vacuum Cutter,Vacuum Mixer
Helper Machinery Group Co., Ltd. , https://www.ihelpergroup.com